Problem:
You can't start the GUI
Solution:
Type kill_busca before you start the GUI. This program deletes all remaining software parts from a previous session.
Problem:
You can't create the focus sequence image automatically
Solution:
Write StopTeleserver in a xterm-window and set RESET tecs-dacs-tvg in the workspace of the tvg.
Problem:
BUSCA GUI is hung
Solution:
Type kill_busca and then start_busca.
Problem:
central 1k+OS bin2 or central 1k bin2 CCD frame definition shows a displacement in some channel or in all of them
Solution:
Reload the CCD database from the menu: CCD->Reload database. There is an opcion from the menu, where is posible to change this database file.
Problem:
You get bias frames, instead of science frames, in the channel a, b and c, but in channel d you can get a science frame
Solution:
Reload the CCD database from the menu: CCD->Reload database. There is an opcion from the menu, where is posible to change this database file.
Problem:
When the TV-Guider is started, you could read: "Electronics initialization failed! caused a problem ... Switch TV-Guider Electronics off/on again and call the staff", and later a "Camera in Overload State" message appear.
Solution:
Don't, worry! It is a normal message, because a BUSCA TV-Guider is being used instead of the normal camera.