Red error flag "AG FOCUS ERROR" in PMAS Status Window.
-
The VME motor controller has experienced an error condition of the A&G Camera
focus stage. The error may be related to a hardware problem, a software
problem, or to a spurious error condition. If the error is prevailing despite
the solution advised below, a more serious hardware failure must be suspected.
Consult support for advice.
agfocReset
limevalSet
agfocInit
agfocStatus
After agfocStatus, the following message should appear: TBD.
If the LOCKED status is not reached, reset the vme controller lea
(reset button on lea frontpanel - do not confuse with abort button !)
Repeat the sequence from above. If there is no success, call support.
In the event of serious failure of the A&G camera focus drive, switch
to the standard 3.5m TV Guider system:
pics_gui pulldown menu: SetMode -> Guiding -> External TV Guider
Note: as opposed to other PMAS devices, you cannot mechanically move
the AG focus stage in case of failure. In such a case it is necessary
to dismount the entire A&G Camera subsystem.