Red error flag "COLPOS ERROR" in PMAS Status Window.
-
The VME motor controller has experienced an error condition.
The error may be related to a hardware problem, a software
problem , or to a spurious error condition. If the error is prevailing despite
the solution advised below, a more serious hardware failure must be suspected.
Consult support for advice.
colReset
limevalSet
colInit
colStatus
After colStatus, the following message should appear:
COL_STATUS: THEVERYFIRST
steps
COL_SET_POSITION: 0
COL_LIMITS: -1600 1100
COL_ACTUAL_POSITION: 0
If this is not the case, reset the vme controller lea (reset
button on lea frontpanel - do not confuse with abort button !)
Repeat the sequence from above. If there is no success, call
support.
Note: it is possible to circumvent software control of this device
to rescue an observation by manually moving the drive into best
focus. There is a vernier screw mounted to the axis of the motor
which can be used for this purpose. FOR CAHA STAFF ONLY !