This Section explains how to perform spectrograph
and A&G camera calibrations
Spectrograph BIAS.
From within pics_GUI, start Bias exposure ()
Spectrograph DARK.
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
dark exposure time
start Dark exposure ()
Spectrograph internal flatfield exposure.
Make sure that CALUNIT is in calibration position
(Status Window CALIBRATION UNIT: CALUNIT Status = O.K. (green).
If not active, select pics_GUI ,
From pics_GUI, select one or several from set of
Calibration Lamps (Lamp 1 ... Lamp 6).
NOTE: up to four lamps can be combined in parallel.
Lamps 2/3 and 4/5 must be selected alternatively since they
are sharing power supplies PS2 and PS3.
pics_GUI: switch on lamp power supplies where needed
pics_GUI: open calibration lamp shutter(s)
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
exposure time
Status Window: make sure calibration lamp LIGHT is
reported on for all selected lamps.
pics_GUI: start exposure ()
Spectrograph sky flatfield exposure.
Make sure that CALUNIT is in park position
(Status Window CALIBRATION UNIT: CALUNIT Status = OFF (red).
If active, select pics_GUI ,
pics_GUI: start exposure ()
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
exposure time
Status Window: make sure calibration lamp LIGHT is
reported on for all selected lamps.
pics_GUI: start exposure ()
A&G Camera BIAS.
On IDL prompt, enter:
pics_a_bias,e
Spectrograph DARK.
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
dark exposure time
start Dark exposure ()
Spectrograph internal flatfield exposure.
Make sure that CALUNIT is in calibration position
(Status Window CALIBRATION UNIT: CALUNIT Status = O.K. (green).
If not active, select pics_GUI ,
From pics_GUI, select one or several from set of
Calibration Lamps (Lamp 1 ... Lamp 6).
NOTE: up to four lamps can be combined in parallel.
Lamps 2/3 and 4/5 must be selected alternatively since they
are sharing power supplies PS2 and PS3.
pics_GUI: switch on lamp power supplies where needed
pics_GUI: open calibration lamp shutter(s)
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
exposure time
Status Window: make sure calibration lamp LIGHT is
reported on for all selected lamps.
pics_GUI: start exposure ()
Spectrograph sky flatfield exposure.
Make sure that CALUNIT is in park position
(Status Window CALIBRATION UNIT: CALUNIT Status = OFF (red).
If active, select pics_GUI ,
pics_GUI: start exposure ()
In pics_GUI, select pulldown menu DefObs - Define PBlock, enter
exposure time
Status Window: make sure calibration lamp LIGHT is
reported on for all selected lamps.